B-scan ultrasound plays an important role in the diagnosis and management of various intraocular disorders in both clear and opaque ocular media. Although B-scan has been considered a useful tool in ophthalmology for several decades, continual developments in ultrasound technology have enhanced the ability of practitioners to detect and differentiate an array of intraocular lesions.

The detection of posterior vitreous detachment (PVD) in opaque ocular media (eg, vitreous hemorrhage) is a well-known indication for B-scan ultrasound. In recent years, however, there has been considerable interest in using B-scan to demonstrate PVD in clear ocular media, in part to assess the efficacy of intraocular drug delivery for the prevention and/or progression of certain disorders. The identification of PVD in clear media can be quite challenging due to the indiscrete nature of the acoustic interface produced by the posterior hyaloid. For this reason, reliable evaluation of clear vitreous requires the use of prescribed examination techniques and a high quality B-scan ultrasound system such as the Eye Cubed (formerly known as I3, Innovative Imaging) from Ellex (Ellex Medical Lasers Ltd., Adelaide, Australia).

PVD FOR PREVENTING
NEOVASCULARIZATION IN NPDR

To date, there are no US Food and Drug Administration (FDA)-approved pharmaceutical treatments available to prevent the occurrence of tractional retinal detachment in the diabetic patient. Current research and investigations, however, may soon lead to the development of such a therapy. Two current clinical trials, the first utilizing sites in the United States and India (http://clinicaltrials.gov/ct2/show/NCT00664183), and the second in Mexico (http://clinicaltrials.gov/ct2/show/ NCT00908778), are investigating the safety and efficacy of intravitreal carbamide (Vitreosolve; Vitreoretinal Technologies, Inc., Irvine, CA), in patients with nonproliferative diabetic retinopathy (NPDR). The primary goal of both studies is to help preserve vision by inducing PVD for the purpose of preventing the progression of diabetic retinopathy. The administered drug induces vitreal detachment by causing vitreous gel to liquefy and thus to take away the hard gel scaffolding that would otherwise support the growth of new blood vessels in patients with NPDR. It thus prevents neovascularizarion (see above, “Intravitreal Carbamide”). The use of high quality B-scan ultrasound instrumentation in this trial is essential to help determine effectiveness of the therapy.

As Director of the Ultrasound Reading Center in Ashville, NC, it has been my role to manage the ultrasound portion of the intravitreal carbamide studies, including devising the B-scan protocol, providing training at the clinical sites, and grading all of the Bscan images produced at the sites. The ultrasound system selected for this study and used by all sites is the Eye Cubed. This system offers numerous superior features that meet all requirements for B-scan ultrasound in the intravitreal carbamide trials, which are listed below.

· High-resolution scanning. The Eye Cubed offers high-resolution scanning of the posterior segment. Using a probe frequency of 10 MHz, the Eye Cubed system provides an axial resolution of 150 μm and a lateral resolution of 200 µm.

· High-speed image acquisition. Supporting an imageacquisition rate of up to 25 frames per second, the Eye Cubed offers high-speed image sampling to allow the observation and storage of findings in real time.

· Network capabilities. Technological developments have also brought improvements in networking capabilities and the ability to capture, store, retrieve, and export movies and still images efficiently.

· Signal-to-noise ratios and artifact reduction. Advancements in amplifier and probe design have resulted in excellent signal-to-noise ratios to minimize artifact and provide high resolution and sensitivity for imaging fine tissue structures.

The culmination of these characteristics of the Eye Cubed system allows the display of PVD in clear ocular media, including epipapillary glial tissue (ie, the Weiss ring) that may assist in establishing the diagnosis of a total PVD (Figure 1).

STANDARDIZATION
Standardization of the protocol across all sites in the intravitreal carbamide trial ensures uniform and accurate data acquisition. A primary benefit of the Eye Cubed is its ability to capture real-time data (movies) using its fast acquisition rate. The use of a standardized protocol, including a set of high-quality movies, allows the independent review of complete screening examinations of the posterior segment and an accurate assessment of the mobility of vitreous pathology. As a result, B-scan data for the study is nearly as good as being able to directly observe the real-time examination at the clinical sites.

In addition to a wide array of technical features, the Eye Cubed has demonstrated high reliability, and Ellex has provided time-tested quality customer service. Such considerations are important when selecting instrumentation to be used in a multicenter clinical trial, and they affected the decision to choose the Eye Cubed system for the intravitreal carbamide trials. Experience with the Eye Cubed in the intravitreal carbamide trials has been excellent as it has fulfilled the requirements of accuracy, stability, and reliability.

The ultrasound technology currently available in ophthalmology has enhanced our ability to diagnose, monitor, and subsequently manage vitreoretinal pathology. As technological advancements continue to emerge, further improvements in B-scan ultrasound technology can be expected.

Sandra Frazier Byrne has been a practitioner and educator in the field of ophthalmic ultrasound since 1971. She practices ophthalmic ultrasound and directs the Ultrasound Reading Center in Asheville, North Carolina. Ms. Byrne can be reached at sandy@suloinc.com.