Retina Pearls is a column that appears regularly in Retina Today. The purpose of the column is to provide a forum for retinal specialists to share informative and exciting tips or pearls with regard to specific vitreoretinal surgical techniques, diagnostics, or therapeutics. This installment of Retina Pearls addresses spectral-domain (SD) optical coherence tomography (OCT). Srinivas R. Sadda, MD, and Pearse Keane, MD, relate that, while the new SD OCT systems are a vast improvement in imaging technology, we must be aware of the variables among the different technologies and how the SD OCT images are interpreted.
We extend an invitation to readers to submit surgical pearls for publication in Retina Today. Please send submissions for consideration to Dean Eliott (deliott@doheny.org) or Ingrid U. Scott, MD, MPH (iscott@psu.edu). We're looking forward to hearing from you. -Dean Eliott, MD; and Ingrid U. Scott, MD, MPH
In recent years, optical coherence tomography (OCT) has become an essential tool for the diagnosis and management of vitreoretinal disorders. In conventional OCT systems, light reflected from the tissue of interest (ie, the neurosensory retina) is assessed as a function of time, and, hence, these systems are often referred to as time-domain (TD) OCT. Stratus OCT (Carl Zeiss Meditec, Dublin, CA), the most commonly used OCT system worldwide, is based on TD technology. Although Stratus OCT is capable of providing relatively high-resolution images (axial: 8 μm to 10 μm), the requirement for a mobile reference mirror limits its image acquisition speed to 400 A-scans per second, and thus, only sparse coverage of the macular area is possible.
Recently, many of the problems associated with TD OCT have been overcome, at least in part, by the release of the next generation of commercial OCT systems. These systems include: Cirrus HD-OCT (Carl Zeiss Meditec), 3D OCT-1000 (Topcon Medical Systems, Paramus, NJ), Spectralis HRA+OCT (Heidelberg Engineering, Heidelberg, Germany), RTVue-100 (Optovue Corporation, Fremont, CA), Spectral OCT SLO (Opko Instrumentation, Miami, FL), SOCT Copernicus HR (Optopol Technology, Zawiercie, Poland), and 3D SDOCT (Bioptigen, Triangle Park, NC). These systems, termed spectral-domain (SD) or Fourierdomain OCT, are based on the assessment of light reflected from the retina as a function of frequency rather than of time. Using these devices, A-scans can be acquired 50 to 100 times more quickly than with TD technology. In addition to their increased image acquisition speed, these instruments offer improvements in sensitivity and image resolution (axial: 5 μm to 7 μm). Although many problems associated with TD OCT have been ameliorated with SD OCT systems, caution is still required as we become increasingly aware of image interpretation pitfalls specific to SD OCT. A few of these potential pitfalls are discussed below.
CONJUGATE IMAGES
SD OCT relies on a combination of spectral interferometry
and Fourier transformation. Fourier transformation
is a mathematical function that is described as
“Hermitian” (ie, it results in the production of two mirror
images, one real and one inverted, which are symmetrical
around a fixed point). SD OCT instruments
generally display only one of the two possible images,
typically the one in which the inner retina is uppermost.
This has the unfortunate effect of limiting the distance
over which OCT images can be obtained. For example,
if a structure spans a large axial distance, such as in eyes
with increased axial length due to pathologic myopia,
portions of the conjugate image may extend onto the
displayed image and be superimposed (Figure 1).
Structures in the vitreous may also be reflected onto or
beneath the retina—this may often result in the hyaloid
face and/or asteroid bodies appearing both within and
underneath the retina (Figure 2).
SEGMENTATION ERRORS
Stratus OCT uses image processing techniques to
automatically detect the inner and outer retinal boundaries
on OCT B-scans (segmentation) and thus provide
measurements of retinal thickness. Using these techniques,
it is possible to measure retinal thickness at multiple locations and to construct retinal thickness
maps. Caution is required in the use of this quantitative
information, as errors in retinal segmentation are
known to occur, and these errors are often severe. SD
OCT systems also utilize automated image analysis software
that provides retinal thickness measurements. In
theory, the greater axial resolution, speed, and sensitivity
of SD OCT should increase the accuracy of these
quantitative assessments. Improvements in computerimage
processing techniques and segmentation algorithms
should also result in a reduced OCT error rate.
Care is still required, however, as the quality of image
segmentation varies between different SD OCT systems,
and severe segmentation errors still occur, resulting in
erroneous retinal thickness measurements. These errors
commonly occur in the setting of epiretinal membranes
or vitreoretinal traction, where the internal limiting
membrane is misidentified. Errors also frequently occur in scans with low signal strength and in patients with
complex morphologic features such as concurrent subretinal
fluid, subretinal tissue, and pigment epithelium
detachment (PED) (Figure 3).
MOTION ARTIFACTS
Due to the slow speed of TD OCT, accurate acquisition
of OCT B-scans is hindered by the presence of
motion artifacts; these artifacts often manifest as undulations
in the neurosensory retina that may make recognition
of certain morphologic features more difficult
(eg, PEDs). The image analysis software of Stratus OCT
performs automatic alignment of A-scans using imageprocessing
techniques to compensate for these motion
artifacts. This function is not required in SD OCT systems
as the high speed scanning of these instruments
prevents the problem of A-scan misalignment within
OCT B-scans. Motion artifacts (eg, vertical microsaccades),
however, remain a problem for SD OCT when
high sampling density is required; for example, when
generating large 3-D data stacks. This may result in inaccurate retinal thickness measurements and the generation
of unrepresentative abnormal features on retinal
thickness maps (Figure 4). For example, at an acquisition
speed of 25,000 A-scans per second, a raster scan
(which consists of a rectangular pattern of horizontal
line scans that run in parallel across the macula) of the
macula consisting of 512 x 512 A-scans (so-called
isotropic sampling) would require more than 10 seconds to be obtained. Many SD OCT systems address
this problem by sacrificing sampling density along one
axis of the raster scan; however, some commercially
available OCT instruments, such as the Spectralis, also
provide real-time eye motion tracking that theoretically
enables longer acquisition times and higher density data
sets. Eye tracking in patients with poor fixation, however,
may also increase the time required to obtain a
dense 3-D set and ultimately reduce the density of
B-scans that can be practically obtained.
RASTER SCAN DENSITY
The limited A-scan acquisition speed of Stratus OCT
restricts both the number of A-scans that can be used
to construct each OCT B-scan, and the number of OCT
B-scans that can be acquired in succession. The high
speed of SD OCT facilitates significantly greater sampling
of the macular area via the use of raster-scanning
protocols. SD OCT systems commonly use raster-scan
protocols that consist of 128 B-scans, with each B-scan
consisting of 512 A-scans (raster scans). The greater retinal
sampling density of SD raster scans may facilitate
early detection of morphologic changes in disease states
and allow these changes to be followed over time more
accurately. SD OCT scanning protocols are still evolving,
however, and the optimal scan density for maximum
clinical applicability remains unknown. The physician
must also be aware that significant differences in retinal
scanning protocols exist between OCT systems. For
example, the Spectralis utilizes multiple B-scan averaging
to produce extremely high quality retinal images; however,
this improvement in image quality may come at
the expense of retinal scanning density.
OCT FUNDUS IMAGES
Another important feature of SD OCT is the ability to
generate OCT fundus images. These images are generated
from the raster scan (3-D) OCT data by summing
the intensity of pixels in the axial direction, resulting in a
pixel brightness value for each axial scan position. OCT
fundus images show a direct view of the macula in
which the retinal vascular arcades may be clearly visible
and spatially consistent with the vasculature on color
photographic or angiographic images. OCT fundus images are useful in the clinical setting as they may facilitate
registration of any point on an OCT image with a
corresponding point on the retinal surface. OCT fundus
images also permit the acquisition of images in the
same location over time. The interpretation of OCT
fundus images, however, requires caution. Although
these images mimic those of standard fundus photography,
many features seen on color photographs may not
be seen on OCT fundus images. In addition, eye motion
artifacts may create abnormal features (Figure 4) and
may reduce the accuracy of registration in SD OCT systems
that depend on the OCT fundus image for B-scan
alignment (ie, systems without eye tracking).
CONCLUSION
SD OCT systems are likely to supplant TD OCT systems
as the standard of care for retina specialists in the
near future. In the early stages of adoption, however, it
is important for retina specialists to recognize some of
the potential pitfalls associated with the use of these
instruments. ■
Srinivas R. Sadda, MD, is an Associate Professor of Ophthalmology at the Doheny Eye Institute and University of Southern California in Los Angeles. He shares in royalties from intellectual property licensed to Topcon Medical Systems and has served as an advisory board member for Heidelberg Engineering. He has also served as a consultant for Genentech, Inc. He can be contacted at +1 323 442 6522; or via e-mail at: SSadda@doheny.org.
Pearse Keane, MD, is a Senior Research Fellow at the Doheny Image Reading Center in Los Angeles. He has no financial interests to disclose. He can be contacted at +1 323 442 6519; or via e-mail at PKeane@doheny.org.
Dean Eliott, MD, is a Professor of Ophthalmology and Director of Clinical Affairs, Doheny Eye Institute, Keck School of Medicine at USC and is a member of the Retina Today Editorial Board. He may be reached by phone: +1 323 442 6582; fax: +1 323 442 6766; or via e-mail: deliott@doheny.org.
Ingrid U. Scott, MD, MPH, is a Professor of Ophthalmology and Public Health Sciences, Penn State College of Medicine, Department of Ophthalmology, and is a member of the Retina Today Editorial Board. She may be reached by phone: +1 717 531 4662; fax: +1 717 531 5475; or via e-mail: iscott@psu.edu.